Pulled-out locking screw re-screwed spontaneously in anterior cervical decompression and fusion with the zero-profile implant system: A case report.

RATIONALE: The zero-profile, standalone device (Zero-P, Synthes GmbH, Switzerland) has been reported to be an effective and safe treatment method with similar clinical outcomes compared with plate. Instrumental complications concerning Zero-P have been little reported. Considering the rarity, we present this amazing case to share with our spinal surgeons and instrument specialists.

PATIENT CONCERNS: A 46-year-old man patient presented to our hospital with neck and shoulders pain for 23 years, numbness and weak-ness of right hand for 6 months.

DIAGNOSES: Hypoesthesia in the right C6 and C7 roots distribution, myodynamia weakness of the right little finger was detected from physical examination. Two-level anterior cervical decompression and fusion (ACDF) using the Zero-P was performed via a classic right Smith-Robinson approach after induction of general anesthesia. Three months postoperative x-rays showed a good position of the implant. Six months postoperative x-rays showed a locking screw at the segment C6/7 pulled out. The patient was diagnosed as screw pullout after ACDF.

INTERVENTIONS: The patient was treated conservatively with regular follow-up as he was asymptomatic and no evidence of esophageal perforation was detected.

OUTCOMES: The patient was followed again and 24 months postoperative x-rays also showed the pulled-out locking screw had re-screwed spontaneously. The patient was noticed that a revision surgery was needed if symptoms occur. At present bony union is not reached but he is still asymptomatic.

LESSONS: Pulled-out screws re-screwed spontaneously are rare but it does occur. Insertion angle may affect the stability of the Zero-P device, and the repeated micro-motion may be the critical reason of the screw pull-out and re-screwing. The management of screws pull-out after ACDF remains individualized and a revision surgery is not necessary for every patient. Conservative treatment such as orthosis and regular follow-ups may be suitable for some asymptomatic patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app