We have located links that may give you full text access.
UK audit of quantitative thyroid uptake imaging.
Nuclear Medicine Communications 2017 July
AIM: A national audit of quantitative thyroid uptake imaging was conducted by the Nuclear Medicine Software Quality Group of the Institute of Physics and Engineering in Medicine in 2014/2015. The aims of the audit were to measure and assess the variability in thyroid uptake results across the UK and to compare local protocols with British Nuclear Medicine Society (BNMS) guidelines.
PARTICIPANTS AND METHODS: Participants were invited through a combination of emails on a public mailbase and targeted invitations from regional co-ordinators. All participants were given a set of images from which to calculate quantitative measures and a spreadsheet for capturing results. The image data consisted of two sets of 10 anterior thyroid images, half of which were acquired after administration of Tc-pertechnetate and the other half after administration of I-iodide. Images of the administration syringes or thyroid phantoms were also included.
RESULTS: In total, 54 participants responded to the audit. The median number of scans conducted per year was 50. A majority of centres had at least one noncompliance in comparison with BNMS guidelines. Of most concern was the widespread lack of injection-site imaging. Quantitative results showed that both intersite and intrasite variability were low for the Tc dataset. The coefficient of quartile deviation was between 0.03 and 0.13 for measurements of overall percentage uptake. Although the number of returns for the I dataset was smaller, the level of variability between participants was greater (the coefficient of quartile deviation was between 0.17 and 0.25).
CONCLUSION: A UK-wide audit showed that thyroid uptake imaging is still a common test in the UK. It was found that most centres do not adhere to all aspects of the BNMS practice guidelines but that quantitative results are reasonably consistent for Tc-based scans.
PARTICIPANTS AND METHODS: Participants were invited through a combination of emails on a public mailbase and targeted invitations from regional co-ordinators. All participants were given a set of images from which to calculate quantitative measures and a spreadsheet for capturing results. The image data consisted of two sets of 10 anterior thyroid images, half of which were acquired after administration of Tc-pertechnetate and the other half after administration of I-iodide. Images of the administration syringes or thyroid phantoms were also included.
RESULTS: In total, 54 participants responded to the audit. The median number of scans conducted per year was 50. A majority of centres had at least one noncompliance in comparison with BNMS guidelines. Of most concern was the widespread lack of injection-site imaging. Quantitative results showed that both intersite and intrasite variability were low for the Tc dataset. The coefficient of quartile deviation was between 0.03 and 0.13 for measurements of overall percentage uptake. Although the number of returns for the I dataset was smaller, the level of variability between participants was greater (the coefficient of quartile deviation was between 0.17 and 0.25).
CONCLUSION: A UK-wide audit showed that thyroid uptake imaging is still a common test in the UK. It was found that most centres do not adhere to all aspects of the BNMS practice guidelines but that quantitative results are reasonably consistent for Tc-based scans.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app