We have located links that may give you full text access.
Accurate Measurement of Indoor Radon Concentration using a Low-Effective Volume Radon Monitor.
Radiation Protection Dosimetry 2017 December 2
AlphaGUARD is a low-effective volume detector and one of the most popular portable radon monitors which is currently available. This study investigated whether AlphaGUARD can accurately measure the variable indoor radon levels. The consistency of the radon-concentration data obtained by AlphaGUARD is evaluated against simultaneous measurements by two other monitors (each ~10 times more sensitive than AlphaGUARD). When accurately measuring radon concentration with AlphaGUARD, we found that the net counts of the AlphaGUARD were required of at least 500 counts, <25% of the relative percent difference. AlphaGUARD can provide accurate measurements of radon concentration for the world average level (~50 Bq m-3) and the reference level of workplace (1000 Bq m-3), using integrated data over at least 3 h and 10 min, respectively.