Technical Note: Direct measurement of continuous TMR data with a 1D tank and automated couch movements.

Medical Physics 2017 July
PURPOSE: Real-time dynamic control of the linear accelerator, couch, and imaging parameters during radiation delivery was investigated as a novel technique for acquiring tissue maximum ratio (TMR) data.

METHODS: TrueBeam Developer Mode (Varian Medical Systems, Palo Alto, CA, USA) was used to control the linear accelerator using the Extensible Markup Language (XML). A single XML file was used to dynamically manipulate the machine, couch, and imaging parameters during radiation delivery. A TG-51 compliant 1D water tank was placed on the treatment couch, and used to position a detector at isocenter at a depth of 24.5 cm. A depth scan was performed towards the water surface. Via XML control, the treatment couch vertical position was simultaneously lowered at the same rate, maintaining the detector position at isocenter, allowing for the collection of TMR data. To ensure the detector remained at isocenter during the delivery, the in-room camera was used to monitor the detector. Continuous kV fluoroscopic images during 10 test runs further confirmed this result. TMR data at multiple Source to Detector Distances (SDD) and scan speeds were acquired to investigate their impact on the TMR data. Percentage depth dose (PDD) scans (for conversion to TMR) along with traditional discrete TMR data were acquired as a standard for comparison.

RESULTS: More than 99.8% of the measured points had a gamma value (1%/1 mm) < 1 when compared with discrete or PDD converted TMR data. Fluoroscopic images showed that the concurrent couch and tank movements resulted in SDD errors < 1 mm. TMRs acquired at SDDs of 99, 100, and 101 cm showed differences less than 0.004.

CONCLUSION: TrueBeam Developer Mode was used to collect continuous TMR data with the same accuracy as traditionally collected discrete data, but yielded higher sampled resolution and reduced acquisition time. This novel method does not require the modification of any equipment and does not use a 3D tank or reservoir.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app