Verification of the Injection Pressure Reduction Effect Using the Novel Indwelling Needle for Contrast-enhanced CT.

PURPOSE: The purpose of this study was to investigate the injection pressure reduction effect of the novel indwelling needle (Becton, Dickinson and Company).

METHOD: We evaluated the period of 651 patients who underwent dynamic computed tomography. We compared the maximum injection pressure. The contrast medium was administered at 320, 350, and 370 mgI/ml.

RESULT: The maximum injection pressure of the novel indwelling needle in 22 G was decreased 10% compared with SC5 in all contrast media. The maximum injection pressure of BDN in 20 G decreases 8% compared with SC5 at 370 mgI/ml, but there was no reduction at 320 mgI/ml and 350 mgI/ml.

CONCLUSION: Our analysis demonstrated that BDN significantly reduced the injection pressure especially in 22 G and using high concentration contrast medium in 20 G.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app