We have located links that may give you full text access.
Dosimetric impacts of endorectal balloon in CyberKnife stereotactic body radiation therapy (SBRT) for early-stage prostate cancer.
PURPOSE: In SBRT for prostate cancer, higher fractional dose to the rectum is a major toxicity concern due to using smaller PTV margin and hypofractionation. We investigate the dosimetric impact on rectum using endorectal balloon (ERB) in prostate SBRT.
MATERIALS AND METHODS: Twenty prostate cancer patients were included in a retrospective study, ten with ERB and 10 without ERB. Optimized SBRT plans were generated on CyberKnife MultiPlan for 5 × 7.25 Gy to PTV under RTOG-0938 protocol for early-stage prostate cancer. For the rectum and the anterior half rectum, mean dose and percentage of volumes receiving 50%, 80%, 90%, and 100% prescription dose were compared.
RESULTS: Using ERB, mean dose to the rectum was 62 cGy (P = 0.001) lower per fraction, and 50 cGy (P = 0.024) lower per fraction for the anterior half rectum. The average V50% , V80% , V90% , and V100% were lower by 9.9% (P = 0.001), 5.3% (P = 0.0002), 3.4% (P = 0.0002), and 1.2% (P = 0.005) for the rectum, and lower by 10.4% (P = 0.009), 8.3% (P = 0.0004), 5.4% (P = 0.0003), and 2.1% (P = 0.003) for the anterior half rectum.
CONCLUSIONS: Significant reductions of dose to the rectum using ERB were observed. This may lead to improvement of the rectal toxicity profiles in prostate SBRT.
MATERIALS AND METHODS: Twenty prostate cancer patients were included in a retrospective study, ten with ERB and 10 without ERB. Optimized SBRT plans were generated on CyberKnife MultiPlan for 5 × 7.25 Gy to PTV under RTOG-0938 protocol for early-stage prostate cancer. For the rectum and the anterior half rectum, mean dose and percentage of volumes receiving 50%, 80%, 90%, and 100% prescription dose were compared.
RESULTS: Using ERB, mean dose to the rectum was 62 cGy (P = 0.001) lower per fraction, and 50 cGy (P = 0.024) lower per fraction for the anterior half rectum. The average V50% , V80% , V90% , and V100% were lower by 9.9% (P = 0.001), 5.3% (P = 0.0002), 3.4% (P = 0.0002), and 1.2% (P = 0.005) for the rectum, and lower by 10.4% (P = 0.009), 8.3% (P = 0.0004), 5.4% (P = 0.0003), and 2.1% (P = 0.003) for the anterior half rectum.
CONCLUSIONS: Significant reductions of dose to the rectum using ERB were observed. This may lead to improvement of the rectal toxicity profiles in prostate SBRT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app