A new immunochemistry platform for a guideline-compliant cardiac troponin T assay at the point of care: proof of principle.
Clinical Chemistry and Laboratory Medicine : CCLM 2017 October 27
BACKGROUND: A multitude of troponin assays for the point-of-care (POC) have been developed showing a lack of analytical sensitivity and precision. We present a new platform solution for the high-sensitivity detection of cardiac troponin T (cTnT) in a 30 μL whole blood sample with a turnaround time of 11 min.
METHODS: The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents. After the sample application, the assay was automatically processed by separating the cellular blood components via centrifugation, followed by incubation of a defined volume from the generated plasma with the immunoreagents. The fluorescence in the signal zone of a membrane was measured after its washing for the cTnT quantitation.
RESULTS: A calibration curve, measured in whole blood samples spiked with native human cTnT, was generated covering a range up to a concentration of approximately 8300 ng/L. The lower detection limit was determined to be 3.0 ng/L. At a concentration of 14 ng/L, the 99th percentile value from the high-sensitivity cardiac troponin T (hs-cTnT) assay in the Elecsys® system, the imprecision (CV) was 3.8%. A CV profile indicated that the functional sensitivity for a CV <10% was 6.8 ng/L. The assay did not show any significant cross-reaction with human skeletal troponin T. We observed an excellent correlation with the hs-TnT Elecsys® assay for 49 clinical plasma samples (r=0.9744).
CONCLUSIONS: The described technology shows that an analytical performance for a highly sensitive determination of cTnT can be achieved in a POC setting.
METHODS: The immunoassay was completely run in a ready-to-use plastic disposable, a centrifugal microfluidic disc with fully integrated reagents. After the sample application, the assay was automatically processed by separating the cellular blood components via centrifugation, followed by incubation of a defined volume from the generated plasma with the immunoreagents. The fluorescence in the signal zone of a membrane was measured after its washing for the cTnT quantitation.
RESULTS: A calibration curve, measured in whole blood samples spiked with native human cTnT, was generated covering a range up to a concentration of approximately 8300 ng/L. The lower detection limit was determined to be 3.0 ng/L. At a concentration of 14 ng/L, the 99th percentile value from the high-sensitivity cardiac troponin T (hs-cTnT) assay in the Elecsys® system, the imprecision (CV) was 3.8%. A CV profile indicated that the functional sensitivity for a CV <10% was 6.8 ng/L. The assay did not show any significant cross-reaction with human skeletal troponin T. We observed an excellent correlation with the hs-TnT Elecsys® assay for 49 clinical plasma samples (r=0.9744).
CONCLUSIONS: The described technology shows that an analytical performance for a highly sensitive determination of cTnT can be achieved in a POC setting.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app