Analysis of Plant Leaves Using Laser Ablation Inductively Coupled Plasma Optical Emission Spectrometry: Use of Carbon to Compensate for Matrix Effects.
Applied Spectroscopy 2017 April
Direct solid sampling by laser ablation into an inductively coupled plasma synchronous vertical dual view optical emission spectroscope (LA-SVDV-ICP-OES) was used for the elemental analysis of nutrient elements Ca, B, Mn, Mg, K, and Zn and essential (non-metallic) elements P and S in plant materials. The samples were mixed with paraffin as a binder, an approach that provides better cohesion of the particles in the pellets in addition to supplying carbon to serve as an internal standard (atomic line C I 193.027 nm) as a way to compensate for matrix effects, and/or variations in the ablation process. Precision was in the range of 1-8% relative standard deviation (RSD) with limit of detection in the range of 0.4-1 mg/kg-1 and 25-640 mg/kg-1 for metallic and non-metallic elements, respectively.