Validation of the A&D UM-201 device for office blood pressure measurement according to the European Society of Hypertension International Protocol Revision 2010.
Blood Pressure Monitoring 2017 August
The aim of this study was to determine the accuracy of the A&D UM-201 device coupled to several cuffs for different arm sizes for office blood pressure (BP) measurement according to the International Protocol of the European Society of Hypertension. Evaluation was carried out in 33 individuals. The mean age of the individuals was 59.3±13.2 years, systolic BP was 145.4±20.6 mmHg (range: 109-186 mmHg), diastolic BP was 87.3±18.0 mmHg (range: 50-124 mmHg), and arm circumference was 30.4±4.2 cm (range: 23-39 cm). The protocol requirements were followed precisely. The UM-201 monitor passed all requirements, fulfilling the standards of the protocol. On average, the device overestimated systolic BP by 3.0±2.1 mmHg and diastolic BP by 2.6±2.0 mmHg. These data show that the A&D UM-201 device coupled to several cuffs for different ranges of arm circumference fulfilled the requirements for validation by the International Protocol and can be recommended for clinical use in the adult population.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app