We have located links that may give you full text access.
EVALUATION STUDIES
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Proof-of-Concept Evaluation of the SailValve Self-Expanding Deep Venous Valve System in a Porcine Model.
Journal of Endovascular Therapy 2017 June
PURPOSE: To evaluate the SailValve, a new self-expanding deep venous valve concept based on a single polytetrafluoroethylene cusp floating up and down in the bloodstream like a sail, acting as a flow regulator and allowing minimal reflux to reduce thrombogenicity.
METHODS: Both iliac veins of 5 pigs were implanted with SailValve devices; the first animal was an acute pilot experiment to show the feasibility of accurately positioning the SailValve via a femoral access. The other 4 animals were followed for 2 weeks (n=2) or 4 weeks (n=2) under a chronic implantation protocol. Patency and valve function were evaluated directly in all animals using ascending and descending phlebography after device placement and at termination in the chronic implant animals. For reasons of clinical relevance, a regimen of clopidogrel and calcium carbasalate was administered. Histological analysis was performed according to a predefined protocol by an independent pathologist.
RESULTS: Deployment was technically feasible in all 10 iliac veins, and all were patent directly after placement. No perioperative or postoperative complications occurred. Ascending phlebograms in the follow-up animals confirmed the patency of all valves after 2 or 4 weeks. Descending phlebograms showed full function in 5 of 8 valves. Limited reflux was seen in 1 valve (4-week group), and the function in the remaining 2 valves (2-week group) was insufficient because of malpositioning. No macroscopic thrombosis was noted on histology. Histology in the follow-up groups revealed a progressive inflammatory reaction to the valves.
CONCLUSION: This animal study shows the potential of the SailValve concept with sufficient valve function after adequate positioning and no (thrombogenic) occlusions after short-term follow-up. Future research is essential to optimize valve material and long-term patency.
METHODS: Both iliac veins of 5 pigs were implanted with SailValve devices; the first animal was an acute pilot experiment to show the feasibility of accurately positioning the SailValve via a femoral access. The other 4 animals were followed for 2 weeks (n=2) or 4 weeks (n=2) under a chronic implantation protocol. Patency and valve function were evaluated directly in all animals using ascending and descending phlebography after device placement and at termination in the chronic implant animals. For reasons of clinical relevance, a regimen of clopidogrel and calcium carbasalate was administered. Histological analysis was performed according to a predefined protocol by an independent pathologist.
RESULTS: Deployment was technically feasible in all 10 iliac veins, and all were patent directly after placement. No perioperative or postoperative complications occurred. Ascending phlebograms in the follow-up animals confirmed the patency of all valves after 2 or 4 weeks. Descending phlebograms showed full function in 5 of 8 valves. Limited reflux was seen in 1 valve (4-week group), and the function in the remaining 2 valves (2-week group) was insufficient because of malpositioning. No macroscopic thrombosis was noted on histology. Histology in the follow-up groups revealed a progressive inflammatory reaction to the valves.
CONCLUSION: This animal study shows the potential of the SailValve concept with sufficient valve function after adequate positioning and no (thrombogenic) occlusions after short-term follow-up. Future research is essential to optimize valve material and long-term patency.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app