Magnetic Mini-Mover Procedure for pectus excavatum IV: FDA sponsored multicenter trial.
Journal of Pediatric Surgery 2017 June
PURPOSE: The Magnetic Mini-Mover Procedure (3MP) is a minimally invasive treatment for prepubertal patients with pectus excavatum. This multicenter trial sought to supplement safety and efficacy data from an earlier pilot trial.
METHODS: Fifteen patients with pectus excavatum had a titanium-enclosed magnet implanted on the sternum. Externally, patients wore a custom-fitted magnetic brace. Patients were monitored closely for safety. Efficacy was determined by the Haller Index (HI) and satisfaction surveys. After 2 years, the implant was removed.
RESULTS: Mean patient age was 12 years (range 8-14), and mean pretreatment HI was 4.7 (range 3.6-7.4). The device was successfully implanted in all patients. Mean treatment duration was 25 months (range 18-33). Posttreatment chest imaging in 13 patients indicated that HI decreased in 5, remained stable in 2, and increased in 6. Seven out of 15 patients had breakage of the implant's titanium cables because of fatigue fracture. Eight out of 13 patients were satisfied with their chest after treatment.
CONCLUSION: The 3MP is a safe, minimally invasive, outpatient treatment for prepubertal patients with pectus excavatum. However, the magnetic implant design led to frequent device breakage, confounding analysis. The HI indicated mixed efficacy, although surveys indicated most patients perceived a benefit.
STUDY TYPE/LEVEL OF EVIDENCE: Case series, treatment study. Level IV.
METHODS: Fifteen patients with pectus excavatum had a titanium-enclosed magnet implanted on the sternum. Externally, patients wore a custom-fitted magnetic brace. Patients were monitored closely for safety. Efficacy was determined by the Haller Index (HI) and satisfaction surveys. After 2 years, the implant was removed.
RESULTS: Mean patient age was 12 years (range 8-14), and mean pretreatment HI was 4.7 (range 3.6-7.4). The device was successfully implanted in all patients. Mean treatment duration was 25 months (range 18-33). Posttreatment chest imaging in 13 patients indicated that HI decreased in 5, remained stable in 2, and increased in 6. Seven out of 15 patients had breakage of the implant's titanium cables because of fatigue fracture. Eight out of 13 patients were satisfied with their chest after treatment.
CONCLUSION: The 3MP is a safe, minimally invasive, outpatient treatment for prepubertal patients with pectus excavatum. However, the magnetic implant design led to frequent device breakage, confounding analysis. The HI indicated mixed efficacy, although surveys indicated most patients perceived a benefit.
STUDY TYPE/LEVEL OF EVIDENCE: Case series, treatment study. Level IV.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app