LSO and LYSO are today the most common scintillators used in positron emission tomography. Lutetium contains traces of 176 Lu, a radioactive isotope that decays β - with a cascade of γ photons in coincidence. Therefore, Lutetium-based scintillators are characterized by a small natural radiation background. In this paper, we investigate and characterize the 176 Lu radiation background via experiments performed on LSO-based PET scanners. LSO background was measured at different energy windows and different time coincidence windows, and by using shields to alter the original spectrum. The effect of radiation background in particularly count-starved applications, such as 90 Y imaging, is analysed and discussed. Depending on the size of the PET scanner, between 500 and 1000 total random counts per second and between 3 and 5 total true coincidences per second were measured in standard coincidence mode. The LSO background counts in a Siemens mCT in the standard PET energy and time windows are in general negligible in terms of trues, and are comparable to that measured in a BGO scanner of similar size.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app