Moving sum of number of positive patient result as a quality control tool.
Clinical Chemistry and Laboratory Medicine : CCLM 2017 October 27
BACKGROUND: Recently, the total prostate-specific antigen (PSA) assay used in a laboratory had a positive bias of 0.03 μg/L, which went undetected. Consequently, a number of post-prostatectomy patients with previously undetectable PSA concentrations (defined as <0.03 μg/L in that laboratory) were being reported as having detectable PSA, which suggested poorer prognosis according to clinical guidelines.
METHODS: Through numerical simulations, we explored (1) how a small bias may evade the detection of routine quality control (QC) procedures with specific reference to the concentration of the QC material, (2) whether the use of 'average of normals' approach may detect such a small bias, and (3) describe the use of moving sum of number of patient results with detectable PSA as an adjunct QC procedure.
RESULTS: The lowest QC level (0.86 μg/L) available from a commercial kit had poor probability (<10%) of a bias of 0.03 μg/L regardless of QC rule (i.e. 1:2S, 2:2S, 1:3S, 4:1S) used. The average number of patient results affected before error detection (ANPed) was high when using the average of normals approach due to the relatively wide control limits. By contrast, the ANPed was significantly lower for the moving sum of number of patient results with a detectable PSA approach.
CONCLUSIONS: Laboratory practitioners should ensure their QC strategy can detect small but critical bias, and may require supplementation of ultra-low QC levels that are not covered by commercial kits with in-house preparations. The use of moving sum of number of patient results with a detectable result is a helpful adjunct QC tool.
METHODS: Through numerical simulations, we explored (1) how a small bias may evade the detection of routine quality control (QC) procedures with specific reference to the concentration of the QC material, (2) whether the use of 'average of normals' approach may detect such a small bias, and (3) describe the use of moving sum of number of patient results with detectable PSA as an adjunct QC procedure.
RESULTS: The lowest QC level (0.86 μg/L) available from a commercial kit had poor probability (<10%) of a bias of 0.03 μg/L regardless of QC rule (i.e. 1:2S, 2:2S, 1:3S, 4:1S) used. The average number of patient results affected before error detection (ANPed) was high when using the average of normals approach due to the relatively wide control limits. By contrast, the ANPed was significantly lower for the moving sum of number of patient results with a detectable PSA approach.
CONCLUSIONS: Laboratory practitioners should ensure their QC strategy can detect small but critical bias, and may require supplementation of ultra-low QC levels that are not covered by commercial kits with in-house preparations. The use of moving sum of number of patient results with a detectable result is a helpful adjunct QC tool.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app