COMPARATIVE STUDY
JOURNAL ARTICLE
Comparison of in-house biotin-avidin tetanus IgG enzyme-linked-immunosorbent assay (ELISA) with gold standard in vivo mouse neutralization test for the detection of low level antibodies.
Journal of Immunological Methods 2017 June
BACKGROUND: Detection of anti-tetanus antibody levels is necessary for both determination of the immune status of individuals and also for planning preventive measures. ELISA is the preferred test among in vitro tests however it can be affected by the cross reacting antibodies. A previously developed in-house ELISA test was found not reliable for the antibody levels ≤1.0IU/ml. A new method was developed to detect low antibody levels correctly. The aim of the present study was to compare the results of the newly developed in-house biotin-avidin tetanus IgG ELISA test with the in vivo mouse neutralization test, for the antibody levels ≤1.0IU/ml.
METHODS: A total of 54 serum samples with the antibody levels of three different levels, =0.01IU/ml, 0.01-0.1IU/ml, 0.1-1IU/ml, which were detected by in vivo mouse neutralization test were studied by the newly developed in-house biotin-avidin tetanus IgG ELISA test. Test was validated by using five different concentrations (0.01IU/ml, 0.06IU/ml, 0.2IU/ml, 0.5IU/ml, 1.0IU/ml).
RESULTS: A statistically significant correlation (r2 =0.9967 p=0,001) between in vivo mouse neutralization test and in-house biotin-avidin tetanus IgG ELISA test, was observed. For the tested concentrations intra-assay, inter-assay, accuracy, sensitivity, specificity and coefficients of variations were determined as ≤15%.
CONCLUSION: In-house biotin-avidin tetanus IgG ELISA test can be an alternative method to in vivo mouse neutralization method for the detection of levels ≤1.0IU/ml. By using in-house biotin-avidin tetanus IgG ELISA test, individuals with non protective levels, will be reliably detected.
METHODS: A total of 54 serum samples with the antibody levels of three different levels, =0.01IU/ml, 0.01-0.1IU/ml, 0.1-1IU/ml, which were detected by in vivo mouse neutralization test were studied by the newly developed in-house biotin-avidin tetanus IgG ELISA test. Test was validated by using five different concentrations (0.01IU/ml, 0.06IU/ml, 0.2IU/ml, 0.5IU/ml, 1.0IU/ml).
RESULTS: A statistically significant correlation (r2 =0.9967 p=0,001) between in vivo mouse neutralization test and in-house biotin-avidin tetanus IgG ELISA test, was observed. For the tested concentrations intra-assay, inter-assay, accuracy, sensitivity, specificity and coefficients of variations were determined as ≤15%.
CONCLUSION: In-house biotin-avidin tetanus IgG ELISA test can be an alternative method to in vivo mouse neutralization method for the detection of levels ≤1.0IU/ml. By using in-house biotin-avidin tetanus IgG ELISA test, individuals with non protective levels, will be reliably detected.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app