Visual or automated dipstick testing for proteinuria in pregnancy?
Pregnancy Hypertension 2017 January
OBJECTIVES: To compare the Multistix 10SG/visual-read with two automated methods (Multistix 10SG/Clinitek 50 and Chemstrip 10A/Urisys 1100) to detect significant proteinuria among high-risk pregnant women.
STUDY DESIGN: Prospective cohort study at British Columbia Women's Hospital & Health Centre, Vancouver, Canada.
MAIN OUTCOME MEASURES: Diagnostic accuracy determined by sensitivity, specificity, positive and negative likelihood ratios (LR+ and LR-).
RESULTS: 303 (89.6%) of 338 women had a urine sample tested by all three dipstick methods. 196 samples (64.7%) were collected in the morning (subsequent to their first void) and from outpatients. 107 samples (35.3%) were from inpatients at various times throughout the day. A PrCr ⩾30mg/mmol was present in 46 (15.2%) samples. The sensitivity for proteinuria was higher with Multistix 10SG/Clinitek 50 (65.2%) than with Multistix 10SG/visual-read (41.3%, p<0.001) or Chemstrip 10A/Urisys 1100 (54.3%, p=0.06). Specificity was >90% for all methods studied, although it was highest for Multistix 10SG/visual-read (98.4%) compared with either Multistix 10SG/Clinitek 50 (92.6%, p<0.001) or Chemstrip 10A/Urisys 1100 (95.7%, p=0.04). For all methods, LR+ was good-excellent (>5), but LR- poor-fair (>0.20). 29 samples were discordant for proteinuria between methods. 28/29 women had negative proteinuria by Multistix 10SG/visual-read, but at least 1+ proteinuria by an automated method; 17/28 were false positives and 11/28 true positives.
CONCLUSIONS: Automated dipstick methods are more sensitive than visual urinalysis for proteinuria, but test performance is still only poor-fair as a 'rule-out' test for proteinuria. Whether the enhanced sensitivity would be worth the false positives, cost, and personnel training remains to be determined for detection of low-level proteinuria in pregnancy.
STUDY DESIGN: Prospective cohort study at British Columbia Women's Hospital & Health Centre, Vancouver, Canada.
MAIN OUTCOME MEASURES: Diagnostic accuracy determined by sensitivity, specificity, positive and negative likelihood ratios (LR+ and LR-).
RESULTS: 303 (89.6%) of 338 women had a urine sample tested by all three dipstick methods. 196 samples (64.7%) were collected in the morning (subsequent to their first void) and from outpatients. 107 samples (35.3%) were from inpatients at various times throughout the day. A PrCr ⩾30mg/mmol was present in 46 (15.2%) samples. The sensitivity for proteinuria was higher with Multistix 10SG/Clinitek 50 (65.2%) than with Multistix 10SG/visual-read (41.3%, p<0.001) or Chemstrip 10A/Urisys 1100 (54.3%, p=0.06). Specificity was >90% for all methods studied, although it was highest for Multistix 10SG/visual-read (98.4%) compared with either Multistix 10SG/Clinitek 50 (92.6%, p<0.001) or Chemstrip 10A/Urisys 1100 (95.7%, p=0.04). For all methods, LR+ was good-excellent (>5), but LR- poor-fair (>0.20). 29 samples were discordant for proteinuria between methods. 28/29 women had negative proteinuria by Multistix 10SG/visual-read, but at least 1+ proteinuria by an automated method; 17/28 were false positives and 11/28 true positives.
CONCLUSIONS: Automated dipstick methods are more sensitive than visual urinalysis for proteinuria, but test performance is still only poor-fair as a 'rule-out' test for proteinuria. Whether the enhanced sensitivity would be worth the false positives, cost, and personnel training remains to be determined for detection of low-level proteinuria in pregnancy.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app