We have located links that may give you full text access.
Case Reports
Journal Article
SafeR and escape junctional rhythm: A singular trigger for pacemaker-mediated tachycardia.
Journal of Electrocardiology 2017 July
A 70-year-old man, who previously received a dual chamber pacemaker for paroxysmal AV block (Sorin Symphony DR 2550), was noted on telemetry to have multiple episodes of rapid ventricular pacing at approximately 120bpm. Evaluation of the telemetry strips revealed that all of the rapid ventricular pacing episodes were initiated by brief runs of escape junctional rhythm. Programmed bradycardia parameters were AAI SafeR with lower rate limit of 50bpm.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.