Fenestration closure with Amplatzer Duct Occluder II in patients after total cavo-pulmonary connection.
Archives of Medical Science : AMS 2017 March 2
INTRODUCTION: Creation of a fenestration during completion of a total cavopulmonary connection (TCPC) has been associated with a reduction in early mortality and morbidity. However, the long-term benefits are negated by an associated limitation in exercise tolerance and the potential risks of thrombo-embolic complications. We sought to describe the safety and efficacy of an Amplatzer Duct Occluder II (ADO II) for transcatheter fenestration closure following TCPC.
MATERIAL AND METHODS: Between January 2000 and July 2014, 102 patients underwent percutaneous closure of extra-cardiac TCPC fenestrations with a range of devices. Patients in whom fenestration closure was performed with an ADO II and who had at least 6 months of follow-up were included in this study.
RESULTS: Forty-seven patients had successful fenestration occlusion with an ADO II. The mean oxygen saturation and mean systemic venous pressures increased from 84.8 ±6.1% before to 97.6 ±2.9% (p < 0.001) after and from 14.2 ±2.15 mm Hg before to 15.6 ±2.2 mm Hg after closure (p < 0.001). Eight patients developed heart failure symptoms, managed by optimization of medical therapy, with 1 patient requiring device removal to reopen the fenestration. Color Doppler transthoracic echocardiography demonstrated residual flow across the device in 18 (38%), 10 (22%), 5 (11%) and 4 (9%) patients before discharge, at 1 and 6 months, and at the latest outpatient visit, respectively.
CONCLUSIONS: The ADO II can be safely and effectively used to close fenestrations in extra-cardiac type Fontan completions. Many of the design features of this device confer potential benefit in this population.
MATERIAL AND METHODS: Between January 2000 and July 2014, 102 patients underwent percutaneous closure of extra-cardiac TCPC fenestrations with a range of devices. Patients in whom fenestration closure was performed with an ADO II and who had at least 6 months of follow-up were included in this study.
RESULTS: Forty-seven patients had successful fenestration occlusion with an ADO II. The mean oxygen saturation and mean systemic venous pressures increased from 84.8 ±6.1% before to 97.6 ±2.9% (p < 0.001) after and from 14.2 ±2.15 mm Hg before to 15.6 ±2.2 mm Hg after closure (p < 0.001). Eight patients developed heart failure symptoms, managed by optimization of medical therapy, with 1 patient requiring device removal to reopen the fenestration. Color Doppler transthoracic echocardiography demonstrated residual flow across the device in 18 (38%), 10 (22%), 5 (11%) and 4 (9%) patients before discharge, at 1 and 6 months, and at the latest outpatient visit, respectively.
CONCLUSIONS: The ADO II can be safely and effectively used to close fenestrations in extra-cardiac type Fontan completions. Many of the design features of this device confer potential benefit in this population.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app