Corneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas.
AIM: To analyze the correlations between ocular biomechanical and biometric data of the eye, measured by Scheimpflug-based devices on healthy subjects.
METHODS: Three consecutive measurements were carried out using the corneal visualization Scheimpflug technology (CorVis ST) device on healthy eyes and the 10 device-specific parameters were recorded. Pentacam HR-derived parameters (corneal curvature radii on the anterior and posterior surfaces; apical pachymetry; corneal volume; corneal aberration data; depth, volume and angle of the anterior chamber) and axial length (AL) from IOLMaster were correlated with the 10 specific CorVis ST parameters.
RESULTS: Measurements were conducted in 43 eyes of 43 volunteers (age 61.24±15.72y). The 10 specific CorVis ST data showed significant relationships with corneal curvature radii both on the anterior and posterior surface, pachymetric data, root mean square (RMS) data of lower-order aberrations, and posterior RMS of higher-order aberrations and spherical aberration of the posterior cornea. Anterior chamber depth showed a significant relationship, but there were no significant correlations between corneal volume, anterior chamber volume, mean chamber angle or AL and the 10 specific CorVis ST parameters.
CONCLUSIONS: CorVis ST-generated parameters are influenced by corneal curvature radii, some corneal RMS data, but corneal volume, anterior chamber volume, chamber angle and AL have no correlation with the biomechanical parameters. The parameters measured by CorVis ST seem to refer mostly to corneal properties of the eye.
METHODS: Three consecutive measurements were carried out using the corneal visualization Scheimpflug technology (CorVis ST) device on healthy eyes and the 10 device-specific parameters were recorded. Pentacam HR-derived parameters (corneal curvature radii on the anterior and posterior surfaces; apical pachymetry; corneal volume; corneal aberration data; depth, volume and angle of the anterior chamber) and axial length (AL) from IOLMaster were correlated with the 10 specific CorVis ST parameters.
RESULTS: Measurements were conducted in 43 eyes of 43 volunteers (age 61.24±15.72y). The 10 specific CorVis ST data showed significant relationships with corneal curvature radii both on the anterior and posterior surface, pachymetric data, root mean square (RMS) data of lower-order aberrations, and posterior RMS of higher-order aberrations and spherical aberration of the posterior cornea. Anterior chamber depth showed a significant relationship, but there were no significant correlations between corneal volume, anterior chamber volume, mean chamber angle or AL and the 10 specific CorVis ST parameters.
CONCLUSIONS: CorVis ST-generated parameters are influenced by corneal curvature radii, some corneal RMS data, but corneal volume, anterior chamber volume, chamber angle and AL have no correlation with the biomechanical parameters. The parameters measured by CorVis ST seem to refer mostly to corneal properties of the eye.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app