Add like
Add dislike
Add to saved papers

Development of a method for rapid analysis of Ra-226 in groundwater and discharge water samples by ICP-QQQ-MS.

A novel and rapid method has been developed for pre-concentration and measurement of226 Ra in groundwater and discharge water samples using the latest generation triple quadrupole inductively coupled plasma mass spectrometry (ICP-QQQ-MS). Cation exchange and extraction chromatography are capable of pre-concentration factors of ~200 based on 1L samples. The sensitivity and interference removal capability of ICP-QQQ-MS was assessed from spiked groundwaters, with the introduction of He collision gas required to minimise instrument background in high-matrix samples. The technique developed is potentially capable of detecting226 Ra activities as low as 5mBqL-1 when combined with pre-concentration prior to measurement.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app