CLINICAL TRIAL
JOURNAL ARTICLE
RESEARCH SUPPORT, NON-U.S. GOV'T
Add like
Add dislike
Add to saved papers

The impact of bowl size, program setup, and blood hematocrit on the performance of a discontinuous autotransfusion system.

Transfusion 2017 March
BACKGROUND: Cell salvage is an essential element in the concept of blood management. Modern devices provide different bowl sizes and sensor-directed programs to optimally adjust to varying clinical situations.

STUDY DESIGN AND METHODS: In an experimental performance study, the discontinuous autotransfusion device XTRA (LivaNova/Sorin) was evaluated using fresh donor blood anticoagulated with heparin 5 U/mL and adjusted to a hematocrit of 10% or 25%, representing orthopedic or cardiac surgery. Test blood was processed with the autotransfusion device XTRA in four different bowls (55 mL, 125 mL, 175 mL, and 225 mL) and in three different program modes (a standard program, an optimized program, and an emergency program).

RESULTS: Processing speed increased with bowl size and with the emergency program (range, 6.4-29.8 mL red blood cells [RBCs]/min). The RBC recovery rate exceeded 90% for all bowls and programs except the 55-mL bowl with the emergency program. Plasma elimination exceeded 95% for all bowls and programs except the 225-mL bowl with the emergency and standard programs. Maximal RBC recovery (range, 94.7%-97.6%) and plasma elimination (range, 98.7%-99.5%) were obtained with the medium-sized bowls (125 mL and 175 mL) and the optimized program. Elimination rates for potassium or plasma free hemoglobin were consistently lower than for protein or albumin and were highest for heparin.

CONCLUSIONS: Increased hematocrit and RBC recovery rates are obtained with the optimized program Popt with the discontinuous autotransfusion device. The emergency program Pem speeds up the process but leads to RBC loss and reduced plasma elimination rates; therefore, it should be restricted to emergency situations. All four different sized bowls have high performance. Plasma elimination is represented best by protein or albumin elimination rates.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app