Multipoint temperature monitoring in liver undergoing computed tomography-guided radiofrequency ablation with fiber Bragg grating probes.

In this work, we investigated the temperature increment experienced by biological tissue during radiofrequency ablation (RFA). The measurements were performed by using two custom-made thermal probes based on fiber optic sensors (fiber Bragg gratings, FBGs). The two probes embed a total of 9 FBGs. Experiments were performed during RFA of an ex vivo healthy porcine liver. The RFA heating module was equipped with 5 thermocouples. Results show that the temperature increment close to the applicator (i.e., 0.6 cm-0.7 cm) reaches the temperature which is set as a target on the RFA module (i.e., approximately 100 °C). The distance from the applicator also has an impact on the dynamics of the heating phenomenon: at short distances the tissue temperature reaches a steady state condition after a few minutes, on the other hand the sensors placed at a distance ≥2cm did not reach the steady-state conditions during the 14-minute procedure. The multipoint temperature monitoring, which uses sensors at several distances from the applicator, can provide useful information regarding the boundary of damaged volume. This approach can be combined with the monitoring temperature system embedded in the heating equipment, to better control the damaged volume, and to improve the treatment outcomes.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app