Add like
Add dislike
Add to saved papers

In vitro assessment of long-term reliability of low-cost μECoG arrays.

Micro-electrocorticographic (μECοG) electrode arrays provide a minimally invasive, high-resolution neural interface with broad cortical coverage. Previously, we fabricated μECoG arrays at a lower cost than commercially available devices using low-cost industrial processes [1], [2]. Here, we report the in vitro electrical performance of five μECoG designs undergoing an accelerated aging protocol. The impedance and yield of the μECoG arrays were tracked over time. The equivalent lifetime at 37°C depended on the manufacturer and material stack-up, and ranged between 30 and greater than 760 days (ongoing). The main failure modes of these devices were delamination at the site of the electrode contact and broken traces due to metal dissolution. Based on these in vitro results, we offer several recommendations for μECoG designs suitable for chronic implantation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app