Add like
Add dislike
Add to saved papers

Analysis and experimental simulation of blood pressure signal distortions observed on intensive care patients monitored by a catheter in radial artery.

The follow-up of patients with hemodynamic instability in intensive care units most often requires blood pressure measurement using a fluid-filled catheter in the radial artery and a fluid-filled tubing connected to a pressure transducer. However, in this usual setup major distortions frequently occur that may alter the pressure signal. Underdamping and overdamping have been well described whereas other types of distortions, including attenuations and varied envelopes of pulse pressure, have been less studied. This study proposes 1) a classification of signal distortions observed on intensive care patients, and 2) several experimental procedures modifying the catheter lumen and the fluid-filled tubing in order to generate a large variety of pressure distortions to more closely mimic the clinical observations.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app