We have located links that may give you full text access.
Medium term review of the ASR implant system: A single surgeon series.
Journal of Orthopaedics 2017 June
INTRODUCTION: Both ASR hip resurfacings and stemmed ASR XL arthroplasties have failed at high rates in several published series. We assessed a single surgeon series of these arthroplasties looking to identify factors associated with their failure.
METHODS: All surgeries were performed by one surgeon. Patients were evaluated clinically, radiologically and with serial cobalt and chromium ion analysis.
RESULTS: 274 implants were analysed - 152 ASR resurfacings and 122 ASR XL implants. Thirty revisions were performed.
CONCLUSION: The failure rate of the ASR implant in our series is unacceptably high - its use in routine hip arthroplasty cannot be supported.
METHODS: All surgeries were performed by one surgeon. Patients were evaluated clinically, radiologically and with serial cobalt and chromium ion analysis.
RESULTS: 274 implants were analysed - 152 ASR resurfacings and 122 ASR XL implants. Thirty revisions were performed.
CONCLUSION: The failure rate of the ASR implant in our series is unacceptably high - its use in routine hip arthroplasty cannot be supported.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app