Tibial Lengthening Using a Fixator-Assisted Lengthening Plate: A New Technique.

Trauma Monthly 2016 November
BACKGROUND: There are many techniques that are used for limb lengthening. Lengthening a limb over a plate is an alternative choice used in children or when using an intramedullary nail is difficult.

OBJECTIVES: In this study, we presented a new technique for tibial lengthening using a monolateral external fixator over a lengthening plate.

MATERIALS AND METHODS: For tibial lengthening, a monolateral external fixator was attached to the composite bone model medially. After a corticotomy was performed, the lengthening plate was placed laterally. Three locking screws were inserted proximally, and two cortical screws were inserted into a lengthening hole that was 1 cm below the osteotomy site. We avoided contact between the screws of the lengthening plate and the pins of the external fixator. During bone lengthening with the monolateral external fixator, the screws at the lengthening hole were able to slide distally with the distal segment of the tibia to allow for tibial elongation. Two locking screws were fixed at the distal locking holes of the plate when the bone elongation was complete. The external fixator was then removed.

RESULTS: The fixator-assisted lengthening plate allowed bone lengthening without malalignment. There were no mechanical problems associated with the external fixator during the lengthening process. Plate osteosynthesis was stable after the fixator was removed. There was no contact between the screws of plate and the Schanz pins of the external fixator under C-arm fluoroscopy.

CONCLUSIONS: The fixator-assisted lengthening plate technique helps to maintain the stability and alignment at both sides of an osteotomy during tibial elongation. It allows the early removal of the external fixator immediately after lengthening is completed. This technique can be applied in children with open physes and in patients with a narrow medullary canal who are unsuitable for limb lengthening over an intramedullary nail.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app