The Interference Phenomenon of Microcatheters in the Jailing Treatment for Internal Carotid Artery Side Wall Aneurysms with an Open Cell Stent System.

OBJECTIVE: Excelsior XT-27 (Stryker Neurovascular, Fremont, CA, USA) or Rebar 27 (eV3 Covidien, Irvine, CA, USA) microcatheters have recently been used to overcome the limitations of the Renegade Hi-Flo microcatheter such as interference between two microcatheters (one for stent delivery and the other for cerebral aneurysm coiling) during the jailing technique. We evaluated differences and influential factors related to the interference phenomenon according to these two microcatheters group.

MATERIALS AND METHODS: Between June 2011 and September 2013, the jailing technique was applied to 94 internal cerebral artery (ICA) aneurysms. The jailing technique with the Neuroform EZ stent system was performed using Renegade (n = 22), Rebar (n = 35), and XT-27 microcatheters (n = 37). In the Renegade Hi-Flo microcatheter group, the jailing technique was successful in 19/22 patients (86.4%) and interference between the two microcatheters occurred in 6/21 patients (28.6%). In the Rebar and XT-27 microcatheter group, the jailing technique was successful in 71/72 patients (98.6%) and interference between the two microcatheters occurred in 1/72 patients (1.4%).

RESULTS: There was a significant difference in the interference between the two delivered microcatheters group (p-value < 0.000) and the carotid siphon angle (p-value: 0.004) in the univariate analysis. In the multiple logistic regression analysis, the Rebar and XT-27 microcatheter group (odds ratio [OD] [95% confidence interval (CI)]; 31.277 [3.138-311.729], p-value: 0.003) and the carotid siphon angle (OD [95%CI]; 0.959 [0.922-0.997], p-value: 0.035) were found to be influential factors in the interference phenomenon.

CONCLUSION: The Rebar 27 and XT-27 microcatheters were more successful and exhibited less interference between the two microcatheters than the Renegade Hi-Flo microcatheter.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app