Iterative Image Reconstruction Improves the Accuracy of Automated Plaque Burden Assessment in Coronary CT Angiography: A Comparison With Intravascular Ultrasound.

OBJECTIVE: The purpose of this study was to determine whether use of iterative image reconstruction algorithms improves the accuracy of coronary CT angiography (CCTA) compared with intravascular ultrasound (IVUS) in semiautomated plaque burden assessment.

MATERIALS AND METHODS: CCTA and IVUS images of seven coronary arteries were acquired ex vivo. CT images were reconstructed with filtered back projection (FBP) and adaptive statistical (ASIR) and model-based (MBIR) iterative reconstruction algorithms. Cross-sectional images of the arteries were coregistered between CCTA and IVUS in 1-mm increments. In CCTA, fully automated (without manual corrections) and semiautomated (allowing manual corrections of vessel wall boundaries) plaque burden assessments were performed for each of the reconstruction algorithms with commercially available software. In IVUS, plaque burden was measured manually. Agreement between CCTA and IVUS was determined with Pearson correlation.

RESULTS: A total of 173 corresponding cross sections were included. The mean plaque burden measured with IVUS was 63.39% ± 10.63%. With CCTA and the fully automated technique, it was 54.90% ± 11.70% with FBP, 53.34% ± 13.11% with ASIR, and 55.35% ± 12.22% with MBIR. With CCTA and the semiautomated technique mean plaque burden was 54.90% ± 11.76%, 53.40% ± 12.85%, 57.09% ± 11.05%. Manual correction of the semiautomated assessments was performed in 39% of all cross sections and improved plaque burden correlation with the IVUS assessment independently of reconstruction algorithm (p < 0.0001). Furthermore, MBIR was superior to FBP and ASIR independently of assessment method (semiautomated, r = 0.59 for FBP, r = 0.52 for ASIR, r = 0.78 for MBIR, all p < 0.001; fully automated, r = 0.40 for FBP, r = 0.37 for ASIR, r = 0.53 for MBIR, all p < 0.001).

CONCLUSION: For the quantification of plaque burden with CCTA, MBIR led to better correlation with IVUS than did traditional reconstruction algorithms such as FBP, independently of the use of a fully automated or semiautomated assessment approach. The highest accuracy for quantifying plaque burden with CCTA can be achieved by using MBIR data with semiautomated assessment.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app