Influences of environmental noise level and respiration rate on the accuracy of acoustic respiration rate monitoring.

We tested the hypothesis that the environmental noise generated by a forced-air warming system reduces the monitoring accuracy of acoustic respiration rate (RRa). Noise levels were adjusted to 45-55, 56-65, 66-75, and 76-85 dB. Healthy participants breathed at set respiration rates (RRset) of 6, 12, and 30/min. Under each noise level at each RRset, the respiration rates by manual counting (RRm) and RRa were recorded. Any appearance of the alarm display on the RRa monitor was also recorded. Each RRm of all participants agreed with each RRset at each noise level. At 45-55 dB noise, the RRa of 13, 17, and 17 participants agreed with RRset of 6, 12, and 30/min, respectively. The RRa of 14, 17, and 16 participants at 56-65 dB noise, agreed with RRset of 6, 12, and 30/min, respectively. At 66-75 dB noise, the RRa of 9, 15, and 16 participants agreed with RRset of 6, 12, and 30/min, respectively. The RRa of one, nine, and nine participants at 76-85 dB noise agreed with RRset of 6, 12, and 30/min, respectively, which was significantly less than the other noise levels (P < 0.05). Overall, 72.9% of alarm displays highlighted incorrect values of RRa. In a noisy situation involving the operation of a forced-air warming system, the acoustic respiration monitoring should be used carefully especially in patients with a low respiration rate.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app