Harmonization of serum 25-hydroxycalciferol assay results from high-performance liquid chromatography, enzyme immunoassay, radioimmunoassay, and immunochemiluminescence systems: A multicenter study.

BACKGROUND: Remarkable disagreement among different systems of 25-hydroxy vitamin D 25(OH)D assay makes decision making for both clinical and community interventions very difficult. This study aimed to harmonize the results obtained from different 25(OH)D assay systems.

METHODS: A total of 275 serum samples were analyzed for 25(OH)D using DIAsource-enzyme immunoassay (EIA), DIAsource-radioimmunoassay (RIA), Roche-electrochemiluminescence (ECL), Diasorin-chemiluminescent immunoassay (CLIA), and high-performance liquid chromatography (HPLC), as the reference method. Serum intact parathyroid hormone (iPTH) was also measured in all samples. Between-system agreement and harmonization were evaluated using Bland-Altman analysis, receiver operating characteristic (ROC), and regression analysis.

RESULTS: Mean serum 25(OH)D concentrations and frequency distribution of vitamin D status showed a significant difference among the studied systems (P<.001 for both). Serum 25(OH)D assay results from all systems correlated with those from HPLC. As compared with HPLC, ECL showed a positive bias (+3.8 nmol/L), whereas CLIA had a negative bias (-11.9 nmol/L). Both EIA and RIA showed a more or less similar positive bias (8.0 and 8.1 nmol/L, respectively). Using serum iPTH-based 25(OH)D cutoff points, only ECL results became comparable to and without significant difference with HPLC. However, when system-specific cutoffs were defined based on HPLC results using regression equations, mean 25(OH)D and frequency distribution of vitamin D status were more harmonized compared with the other methods.

CONCLUSION: Our findings showed that with adjustment of circulating 25(OH)D based on HPLC, frequency distribution of vitamin D status, as judged by different methods, can be well harmonized with no statistically significant inter-system difference.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app