Reproducibility of a Noninvasive System for Eye Positioning and Monitoring in Stereotactic Radiotherapy of Ocular Melanoma.

PURPOSE: Our preferred treatment for juxtapapillary choroidal melanoma is stereotactic radiotherapy. We aim to describe our immobilization system and quantify its reproducibility.

MATERIALS AND METHODS: Patients were identified in our radiosurgery database. Patients were imaged at computed tomography simulator with an in-house system which allows visual monitoring of the eye as the patient fixates a small target. All patients were reimaged at least once prior to and/or during radiotherapy. The patients were treated on the CyberKnife system, 60 Gy in 10 daily fractions, using skull tracking in conjunction with our visual monitoring system. In order to quantify the reproducibility of the eye immobilization system, computed tomography scans were coregistered using rigid 6-dimensional skull registration. Using the coregistered scans, x, y, and z displacements of the lens/optic nerve insertion were measured. From these displacements, 3-dimensional vectors were calculated.

RESULTS: Thirty-four patients were treated from October 2010 to September 2015. Thirty-nine coregistrations were performed using 73 scans (2-3 scans per patient). The mean displacements of lens and optic nerve insertion were 0.1 and 0.0 mm. The median 3-dimensional displacements (absolute value) of lens and nerve insertion were 0.8 and 0.7 mm (standard deviation: 0.5 and 0.6 mm). Ninety-eight percent of 3-dimensional displacements were below 2 mm (maximum 2.4 mm). The calculated planning target volume (PTV) margins were 0.8, 1.4, and 1.5 mm in the anterior-posterior, craniocaudal, and right-left axes, respectively. Following this analysis, no further changes have been applied to our planning margin of 2 to 2.5 mm as it is also meant to account for uncertainties in magnetic resonance imaging to computed tomography registration, skull tracking, and also contouring variability.

CONCLUSION: We have found our stereotactic eye immobilization system to be highly reproducible (<1 mm) and free of systematic error.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app