A three-dimensional visualization preoperative treatment planning system for microwave ablation in liver cancer: a simulated experimental study.
Abdominal Radiology 2017 June
PURPOSE: To evaluate the application value of three-dimensional (3D) visualization preoperative treatment planning system (VPTPS) for microwave ablation (MWA) in liver cancer.
MATERIALS AND METHODS: The study was a simulated experimental study using the CT imaging data of patients in DICOM format in a model. Three students (who learn to interventional ultrasound for less than 1 year) and three experts (who have more than 5 years of experience in ablation techniques) in MWA performed the preoperative planning for 39 lesions (mean diameter 3.75 ± 1.73 cm) of 32 patients using two-dimensional (2D) image planning method and 3D VPTPS, respectively. The number of planning insertions, planning ablation rate, and damage rate to surrounding structures were compared between2D image planning group and 3D VPTPS group.
RESULTS: There were fewer planning insertions, lower ablation rate and higher damage rate to surrounding structures in 2D image planning group than 3D VPTPS group for both students and experts. When using the 2D ultrasound planning method, students could carry out fewer planning insertions and had a lower ablation rate than the experts (p < 0.001). However, there was no significant difference in planning insertions, the ablation rate, and the incidence of damage to the surrounding structures between students and experts using 3D VPTPS.
CONCLUSIONS: 3DVPTPS enables inexperienced physicians to have similar preoperative planning results to experts, and enhances students' preoperative planning capacity, which may improve the therapeutic efficacy and reduce the complication of MWA.
MATERIALS AND METHODS: The study was a simulated experimental study using the CT imaging data of patients in DICOM format in a model. Three students (who learn to interventional ultrasound for less than 1 year) and three experts (who have more than 5 years of experience in ablation techniques) in MWA performed the preoperative planning for 39 lesions (mean diameter 3.75 ± 1.73 cm) of 32 patients using two-dimensional (2D) image planning method and 3D VPTPS, respectively. The number of planning insertions, planning ablation rate, and damage rate to surrounding structures were compared between2D image planning group and 3D VPTPS group.
RESULTS: There were fewer planning insertions, lower ablation rate and higher damage rate to surrounding structures in 2D image planning group than 3D VPTPS group for both students and experts. When using the 2D ultrasound planning method, students could carry out fewer planning insertions and had a lower ablation rate than the experts (p < 0.001). However, there was no significant difference in planning insertions, the ablation rate, and the incidence of damage to the surrounding structures between students and experts using 3D VPTPS.
CONCLUSIONS: 3DVPTPS enables inexperienced physicians to have similar preoperative planning results to experts, and enhances students' preoperative planning capacity, which may improve the therapeutic efficacy and reduce the complication of MWA.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app