A Novel Automated Somatosensory Evoked Potential (SSEP) Monitoring Device for Detection of Intraoperative Peripheral Nerve Injury in Cardiac Surgery: A Clinical Feasibility Study.

OBJECTIVE: Peripheral nerve injury (PNI) is a common and potentially devastating complication in cardiac surgery. Somatosensory evoked potential (SSEP) monitoring is one of the modalities for PNI; however, its application is limited by complicated logistics. This study aimed to assess the feasibility of using a novel, automated SSEP device (EPAD; SafeOp Surgical, Hunt Valley, MD) for detection of intraoperative PNI during cardiac surgery.

DESIGN: Prospective, observational study.

SETTING: Single university hospital.

PARTICIPANTS: Cardiac surgical patients.

INTERVENTIONS: After Ethics Board approval and written consent, study participants were monitored using the EPAD automated SSEP device during cardiac surgery. All patients with prolonged and abnormal SSEP changes were evaluated postoperatively, and if they were symptomatic, they were referred for further nerve conduction and electromyographic assessment.

MEASUREMENTS AND MAIN RESULTS: Of the 43 patients who consented to study inclusion, 33 were monitored successfully. With increasing clinical experience the authors encountered minimal technical issues, and satisfactory signals were obtained in most patients. Abnormal SSEP signal changes, which were encountered in 5 (15.2%) patients, were interpreted as impending PNI; 3 patients experienced prolonged signal changes (>1 h), and 2 (6.1%) of these developed symptomatic peripheral neuropathy that was confirmed with nerve conduction studies.

CONCLUSIONS: The EPAD automated SSEP device is a viable option for detecting PNI during cardiac surgery. A high incidence of intraoperative peripheral nerve compromise and a 6.1% incidence of postoperative peripheral neuropathy were observed. This study reports the clinical feasibility of using the EPAD automated SSEP device; additional studies are required to evaluate the diagnostic test accuracy and the outcome benefit of routine SSEP monitoring in cardiac surgical patients.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app