Intramedullary cortical bone strut improves the cyclic stability of osteoporotic proximal humeral fractures.

BACKGROUND: Proximal humeral fractures treated with locking plate can fail due to varus collapse, especially in osteoporotic bone with medial cortex comminution. The use of an intramedullary strut together with locking plate fixation may strengthen fixation and provide additional medial support to prevent the varus malalignment. This study biomechanically investigates the influence of an intramedullary cortical bone strut on the cyclic stability of proximal humeral fractures stabilized by locking plate fixation in a cadaver model.

METHODS: Ten cadaveric humeri were divided into two groups statistically matched for bone density. Each specimen was osteotomized with 10 mm gap at the surgical neck. The non-augmented group stabilized with locking plate alone; in the augmented group, a locking plate was used combined with an intramedullary cortical bone strut. The strut was retrograded into the subchondral bone, and three humeral head screws were inserted into the strut to form a plate-screw-strut mechanism. The cyclic axial load was performed to 450 N for 6000 cycles and then loaded to failure. Construct stiffness, cyclic loading behavior and failure strength were analyzed to identify differences between groups.

RESULTS: The augmented constructs were significantly stiffer than the non-augmented constructs during cycling. On average, the maximum displacements at 6000 cycles for non-augmented and augmented groups were 3.10 ± 0.75 mm and 1.7 ± 0.65 mm (p = 0.01), respectively. The mean peak-to-peak (inter cycle) displacement at 6000 cycles was about 2 times lower for the augmented group (1.36 ± 0.68 mm vs. 2.86 ± 0.51 mm). All specimens showed varus collapse combined with loss of screw fixation of the humeral head. The failure load of the augmented group was increased by 2.0 (SD = 0.41) times compared with the non-augmented group (p < 0.001).

CONCLUSIONS: The stability and strength of the locking plate augmented with an intramedullary strut were significantly increased. For bone with poor quality, the subsidence of the locked screws led larger displacement, decreased the stability of the constructs, however, the plate-screw-strut mechanism provided more rigidity to stabilize the fixation. This study emphasized the importance of intramedullary support for the proximal humeral fractures fixed with a locked plate under cyclic loading, especially in bone with poor quality. This work is based on the results of cadaver model, further in vivo analysis is necessary to determine if the clinical results can be extrapolated from this data.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app