Pressures and Oscillation Frequencies Generated by Bubble-Positive Expiratory Pressure Devices.

Respiratory Care 2017 April
BACKGROUND: Positive expiratory pressure (PEP) devices are used to assist with airway clearance. Little is known about the therapist-made or commercially available bubble-PEP devices. The aim of this study was to determine the end-expiratory pressures (cm H2 O) and oscillation frequencies (Hz) generated when a range of flows were applied to the therapist-made bubble-PEP devices (Bubble-PEP-3cm and Bubble-PEP-0cm) and commercial bubble-PEP devices (AguaPEP, Hydrapep, and Therabubble).

METHODS: This was a bench-top experimental study using a compressed air source, flow rotameter (flows of 5, 10, 15, 20, and 25 L/min), and pressure transducer. Data were collected using a data acquisition device with PhysioDAQxs software and analyzed with Breathalyser software to determine the pressures and oscillation frequencies generated by 5 bubble-PEP devices. Each flow was constant for a 30-s measurement period, and measurements were repeated in triplicate. The 5 devices were: a therapist-made Bubble-PEP-3cm device (filled with 13 cm of water, tubing resting 3 cm from the base of the container); the therapist-made Bubble-PEP-0cm (filled with 10 cm of water, tubing resting at the base of the container); and the AguaPEP, Hydrapep, and Therabubble devices with water to the 10 cm mark on the containers.

RESULTS: Flows of 5-25 L/min produced the following mean ± SD PEP and oscillation frequencies (Hz): the Bubble-PEP-3cm produced PEP of 10.4 ± 0.14 to 10.8 ± 0.24 cm H2 O, oscillations between 13 and 17 Hz; the Bubble-PEP-0cm produced PEP of 10.9 ± 0.01 to 12.9 ± 0.08 cm H2 O, oscillations between 12 and 14 Hz; the AguaPEP produced PEP from 9.7 ± 0.02 to 11.5 ± 0.02 cm H2 O, oscillations between 11 and 17 Hz; the Hydrapep produced PEP of 9.6 ± 0.35 to 10.7 ± 0.39 cm H2 O, oscillations between 14 and 17 Hz; and the Therabubble produced PEP from 8.6 ± 0.01 to 12.8 ± 0.03 cm H2 O, oscillations between 14 and 17 Hz.

CONCLUSIONS: Bubble-PEP-3cm maintained the most stable pressure throughout the range of flows tested. All devices investigated produced similar oscillation frequencies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app