Comparative Study
Journal Article
Add like
Add dislike
Add to saved papers

Computer-based comparison of different methods for selecting mitral annuloplasty ring size.

BACKGROUND: Ring sizing for mitral valve annuloplasty is conventionally done intraoperatively using specific 'sizer' instruments, which are placed onto the valve tissue. This approach is barely reproducible since different sizing strategies have been established among surgeons. The goal of this study is to virtually apply different sizing methods on the basis of pre-repair echocardiography to find out basic differences between sizing strategies.

METHODS: In three-dimensional echocardiographs of 43 patients, the mitral annulus and the contour of the anterior mitral leaflet were segmented using MITK Mitralyzer software. Similarly, three-dimensional virtual models of Carpentier-Edwards Physio II annuloplasty rings and their corresponding sizers were interactively generated from computer tomography images. For each patient, the matching annuloplasty ring was selected repeatedly according to popular sizing strategies, such as the height of anterior mitral leaflet, the intercommissural distance and the surface area of anterior mitral leaflet. The areas of the selected rings were considered as the neo-surface area of the mitral annulus after implantation.

RESULTS: The sizing of the mitral valve according to the height of anterior mitral leaflet (mean ring size = 29.9 ± 3.90), intercommissural distance (mean ring size = 37.5 ± 1.92) or surface area of anterior mitral leaflet (mean ring size = 32.7 ± 3.3) led to significantly different measurements (p ≤ 0.01). In contrary to intercommissural distance, height and surface area of the anterior mitral leaflet exhibited significant variations between the patients (p ≤ 0.01). The sizing according to the height of anterior mitral leaflet led to the maximal reduction of the mitral annulus surface area followed by the sizing according to the surface area of anterior mitral leaflet and finally by the intercommissural distance.

CONCLUSIONS: This novel comprehensive computer-based analysis reveals that the surveyed sizing methods led to the selection of significantly different annuloplasty rings and therefore underscore the ambiguity of routinely applied annuloplasty sizing strategies.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app