Comparative Study
Journal Article
Research Support, Non-U.S. Gov't
Add like
Add dislike
Add to saved papers

Large-scale Retrospective Monte Carlo Dosimetric Study for Permanent Implant Prostate Brachytherapy.

PURPOSE: To retrospectively compare water-based and full tissue model Monte Carlo dose calculations in a large cohort of patients undergoing 125 I permanent implant prostate brachytherapy.

METHODS AND MATERIALS: For 613 patients, EGSnrc BrachyDose dose calculations were performed in 2 virtual patient models: TG43sim (simulated American Association of Physicists in Medicine Task Group Report 43 conditions) and MCref (computed tomography-derived heterogeneous tissue model with interseed effects). A sensitivity analysis was performed in a patient subset (25 with and 25 without prostatic calcifications) to explore dose calculation dependence on organ-at-risk (OAR) and calcification tissue elemental compositions and modelling approach.

RESULTS: In the target volume, the minimum radiation dose delivered to 90% of prostate (D90 ) (volume of prostate receiving at least 100% of prescription dose [V100 ]) was lower with MCref than with TG43sim by 5.9% ± 1.6% (2.6% ± 1.7%), on average. Patients with prostatic calcifications can have substantial underdosed volumes due to calcification shielding, lowering the D90 by ≤25%. In the urethra, the average D5 (D30 ) was lower with MCref than with TG43sim by 4.4% ± 1.8% (4.7% ± 1.9%). In the rectum (bladder), the minimum dose to the hottest 0.1 cm3 (D_0.1cm3 ) of the contoured organ was lower (higher) with MCref than with TG43sim by 5.2% ± 1.8% (1.3% ± 1.8%). Doses to the target and OARs can increase or decrease by several percentages, depending on the assumed tissue elemental composition. In patients with calcifications, differences between approaches to model calcifications can change the target and OAR dose metrics by upward of 10%.

CONCLUSIONS: TG43sim typically overestimates the target and OAR doses by several percentages, on average, compared with MCref. The considerable variation in the relative TG43sim and MCref doses between patients, and the larger dose differences for patients with calcification, suggests that clinical adoption of Monte Carlo dose calculations for permanent implant prostate brachytherapy should be pursued. The substantial sensitivity of the Monte Carlo dose calculations to the patient modelling approach supports the adoption of a consensus modelling scheme, such as MCref described in the present study, to ensure consistency of practice.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app