Closing the gap in paediatric ventricular assist device therapy with the Berlin Heart EXCOR® 15-ml pump†.
Interactive Cardiovascular and Thoracic Surgery 2017 January 24
OBJECTIVES: The Berlin Heart EXCOR(®) (EXCOR) paediatric ventricular assist device is used worldwide for mechanical support of infants and small children with end-stage heart failure. A clinically important gap between the smallest EXCOR blood pump (10 ml) and the next larger size (25 ml) limited the choice of pump size in patients with a body surface area (BSA) between 0.33 and 0.5 m(2) We present the first clinical experience from the early product surveillance (EPS) of the new EXCOR 15-ml blood pump.
METHODS: After CE and U.S. Food and Drug Administration approval in January 2013, 20 patients with a mean age of 1.6 years (range 0.5-3.5 years) and a mean BSA of 0.45 m(2) (range 0.33-0.59 m(2)) were enrolled in the EPS. The main diagnosis was idiopathic cardiomyopathy in 13 patients; the majority (n = 16) of children were in INTERMACS level 1 or 2. Data from high-volume paediatric transplant centres were collected prospectively for a defined follow-up period of 60 days after device implantation.
RESULTS: Mean time on the EXCOR 15-ml blood pump was 43 days; the survival rate was 100% at the end of the EPS period. Seven patients underwent a heart transplant from the device; 2 children were weaned; and 11 patients remained on support. Infection of cannula exit sites occurred in 3 patients. Two patients had minor thromboembolic strokes but made a complete neurological recovery.
CONCLUSIONS: The new EXCOR 15-ml blood pump demonstrated optimal ventricular assist device support of children with a BSA of 0.33-0.5 m(2).
METHODS: After CE and U.S. Food and Drug Administration approval in January 2013, 20 patients with a mean age of 1.6 years (range 0.5-3.5 years) and a mean BSA of 0.45 m(2) (range 0.33-0.59 m(2)) were enrolled in the EPS. The main diagnosis was idiopathic cardiomyopathy in 13 patients; the majority (n = 16) of children were in INTERMACS level 1 or 2. Data from high-volume paediatric transplant centres were collected prospectively for a defined follow-up period of 60 days after device implantation.
RESULTS: Mean time on the EXCOR 15-ml blood pump was 43 days; the survival rate was 100% at the end of the EPS period. Seven patients underwent a heart transplant from the device; 2 children were weaned; and 11 patients remained on support. Infection of cannula exit sites occurred in 3 patients. Two patients had minor thromboembolic strokes but made a complete neurological recovery.
CONCLUSIONS: The new EXCOR 15-ml blood pump demonstrated optimal ventricular assist device support of children with a BSA of 0.33-0.5 m(2).
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app