MRI, Retrieval Analysis, and Histologic Evaluation of Adverse Local Tissue Reaction in Metal-on-Polyethylene Total Hip Arthroplasty.

BACKGROUND: The severity and location of adverse local tissue reaction (ALTR) seen in metal-on-polyethylene total hip arthroplasty (THA) is not well described.

METHODS: We retrospectively reviewed the records of patients who underwent a revision THA using our biomechanics database. We included all patients who underwent revision surgery for the diagnosis of ALTR with THA implants that had modularity solely at the head-neck junction and excluded patients with implant modularity at sites other than the head-neck junction. Magnetic resonance imaging (MRI) was evaluated by a fellowship-trained radiologist who specializes in evaluating metal artifact reducing MRI sequences to quantify the ALTR lesions. Histology was evaluated for findings of ALTR using the Campbell score.

RESULTS: We identified 11 patients in the database. Eight patients had an MRI ALTR grade of severe and 7 did based on the histology score. The mean synovial volume was 218,658 mm3 (range 23,461-451,435 mm3 ) with a mean maximal synovial thickness of 15.3 mm (range 3-34.3 mm). A disruptive infiltration of the abductors due to pseudocapsule invasion was seen in 67% of the patients with 3 not having abductor involvement. Mean preoperative cobalt and chromium levels were 5.4 ppb (range 1-12.3 ppb) and 1.1 ppb (range 0.6-2.4 ppb), respectively.

CONCLUSION: Patients with ALTR from head-neck junction corrosion in metal-on-polyethylene THA may present with large pseudotumors that have previously been under appreciated. Mean preoperative serum cobalt and chromium ion levels remained relatively low, and MRI was an effective way to characterize the size and location of these lesions.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app