The concurrent validity and reliability of the Leg Motion system for measuring ankle dorsiflexion range of motion in older adults.

PeerJ 2017
BACKGROUND: New reliable devices for range of motion (ROM) measures in older adults are necessary to improve knowledge about the functional capability in this population. Dorsiflexion ROM limitation is associated with ankle injuries, foot pain, lower limb disorders, loss of balance, gait control disorders and fall risk in older adults. The aim of the present study was to assess the validity and reliability of the Leg Motion device for measuring ankle dorsiflexion ROM in older adults.

METHODS: Adescriptive repeated-measures study was designed to test the reliability of Leg Motion in thirty-three healthy elderly patients older than 65 years. The subjects had to meet the following inclusion and exclusion criteria in their medical records: older than 65 years; no lower extremity injury for at least one year prior to evaluation (meniscopathy, or fractures) and any chronic injuries (e.g., osteoarthritis); no previous hip, knee or ankle surgery; no neuropathic alterations and no cognitive conditions (e.g., Alzheimer's disease or dementia). Participants were recruited through the person responsible for the physiotherapist area from a nursing center. The subjects were evaluated in two different sessions at the same time of day, and there was a break of two weeks between sessions. To test the validity of the Leg Motion system, the participants were measured in a weight-bearing lunge position using a classic goniometer with 1° increments, a smartphone with an inclinometer standard app (iPhone 5S® ) with 1° increments and a measuring tape that could measure 0.1 cm. All testing was performed while the patients were barefoot. The researcher had ten years of experience as a physiotherapist using goniometer, tape measure and inclinometer devices.

RESULTS: Mean values and standard deviations were as follows: Leg Motion (right 5.15 ± 3.08; left 5.19 ± 2.98), tape measure (right 5.12 ± 3.08; left 5.12 ± 2.80), goniometer (right 45.87° ± 4.98; left 44.50° ± 5.54) and inclinometer app (right 46.53° ± 4.79; left 45.27° ± 5.19). The paired t -test showed no significant differences between the limbs or between the test and re-test values. The test re-test reliability results for Leg Motion were as follows: the standard error of the measurement ranged from 0.29 to 0.43 cm, the minimal detectable difference ranged from 0.79 to 1.19 cm, and the intraclass correlation coefficients (ICC) values ranged from 0.97 to 0.98.

CONCLUSIONS: The results of the present study indicated that the Leg Motion device is a valid, reliable, accessible and portable tool as an alternative to the classic weight-bearing lunge test for measuring ankle dorsiflexion ROM in older adults.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app