Comparison of valve performance of the mechanically expanding Lotus and the balloon-expanded SAPIEN3 transcatheter heart valves: an observational study with independent core laboratory analysis.
European Heart Journal Cardiovascular Imaging 2018 Februrary 2
Aims: The Lotus and SAPIEN3 are second-generation transcatheter heart valves, which are designed to minimize paravalvular aortic regurgitation (PAR) after transcatheter aortic valve replacement (TAVR). We sought to compare both devices for valve performance and with emphasis on PAR by independent core laboratory analysis.
Methods and results: A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 80 ± 8 years, Logistic EuroSCORE 14.8 ± 9.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-treated patients and frequency of post-dilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation (P = 0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P = 0.03) occurrence of PAR.
Conclusion: Overall, haemodynamic performance was comparable between the Lotus and SAPIEN3 valves. Lotus valve required less oversizing and was associated with less PAR than SAPIEN3.
Methods and results: A total of 162 (79 Lotus and 83 SAPIEN3) consecutive patients (51% female, 80 ± 8 years, Logistic EuroSCORE 14.8 ± 9.4%) who underwent TAVR because of aortic stenosis were included. Patients with aortic valve-in-valve treatment were excluded. Pre-discharge echocardiograms were analysed by an independent core laboratory using the Valve Academic Research Consortium 2 criteria. There were no differences in baseline and procedural characteristics, except for a larger aortic annulus and sizing indices in SAPIEN3-treated patients and frequency of post-dilatation (0% in Lotus and 13.1% in SAPIEN3). Both valves have similar mean residual gradient, indexed effective orifice area and Doppler velocity index when adjusted to valve size. The frequency of mild (13.9% vs. 31.3%) and at least moderate (1.3% vs. 3.6%) PAR was less after Lotus than after SAPIEN3 implantation (P = 0.02). Multi-slice computed tomography-based annulus and left ventricular outflow tract diameters, calcification and percentage of oversizing were not different between those with or without mild or more PAR. On multivariate analysis, the use of Lotus valve was associated with less (odds ratio OR, 0.41, P = 0.03) occurrence of PAR.
Conclusion: Overall, haemodynamic performance was comparable between the Lotus and SAPIEN3 valves. Lotus valve required less oversizing and was associated with less PAR than SAPIEN3.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app