SU-F-I-58: Image Quality Comparisons of Different Motion Magnitudes and TR Values in MR-PET.

Medical Physics 2016 June
PURPOSE: The aim of this work is to evaluate the accuracy and sensitivity of a respiratory-triggered MR-PET protocol in detecting four different sized lesions at two different magnitudes of motion, with two different TR values, using a novel PET-MR-CT compatible respiratory motion phantom.

METHODS: The eight-compartment torso phantom was setup adjacent to the motion stage, which moved four spherical compartments (28, 22, 17, 10 mm diameter) in two separate (1 and 2 cm) linear motion profiles, simulating a 3.5 second respiratory cycle. Scans were acquired on a 3T MR-PET system (Biograph mMR; Siemens Medical Solutions, Germany). MR measurements were taken with: 1) Respiratory-triggered T2-weighted turbo spin echo (BLADE) sequence in coronal orientation, and 2) Real-time balanced steady-state gradient echo sequence (TrueFISP) in coronal and sagittal planes. PET was acquired simultaneously with MR. Sphere geometries and motion profiles were measured and compared with ground truths for T2 BLADE-TSE acquisitions and real time TrueFISP images. PET quantification and geometry measurements were taken using standardized uptake values, voxel intensity plots and were compared with known values, and examined alongside MR-based attenuation maps. Contrast and signal-to-noise ratios were also compared for each of the acquisitions as functions of motion range and TR.

RESULTS: Comparison of lesion diameters indicate the respiratory triggered T2 BLADE-TSE was able to maintain geometry within -2 mm for 1 cm motion for both TR values, and within -3.1 mm for TR = 2000 ms at 2 cm motion. Sphere measurements in respiratory triggered PET images were accurate within +/- 5 mm for both ranges of motion for 28, 22, and 17 mm diameter spheres.

CONCLUSION: Hybrid MR-PET systems show promise in imaging lung cancer in non-compliant patients, with their ability to acquire both modalities simultaneously. However, MR-based attenuation maps are still susceptible to motion derived artifacts and pose the potential to affect PET accuracy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app