SU-F-I-43: A Software-Based Statistical Method to Compute Low Contrast Detectability in Computed Tomography Images.

Medical Physics 2016 June
PURPOSE: The low contrast detectability (LCD) of a CT scanner is its ability to detect and display faint lesions. The current approach to quantify LCD is achieved using vendor-specific methods and phantoms, typically by subjectively observing the smallest size object at a contrast level above phantom background. However, this approach does not yield clinically applicable values for LCD. The current study proposes a statistical LCD metric using software tools to not only to assess scanner performance, but also to quantify the key factors affecting LCD. This approach was developed using uniform QC phantoms, and its applicability was then extended under simulated clinical conditions.

METHODS: MATLAB software was developed to compute LCD using a uniform image of a QC phantom. For a given virtual object size, the software randomly samples the image within a selected area, and uses statistical analysis based on Student's t-distribution to compute the LCD as the minimal Hounsfield Unit's that can be distinguished from the background at the 95% confidence level. Its validity was assessed by comparison with the behavior of a known QC phantom under various scan protocols and a tissue-mimicking phantom. The contributions of beam quality and scattered radiation upon the computed LCD were quantified by using various external beam-hardening filters and phantom lengths.

RESULTS: As expected, the LCD was inversely related to object size under all scan conditions. The type of image reconstruction kernel filter and tissue/organ type strongly influenced the background noise characteristics and therefore, the computed LCD for the associated image.

CONCLUSION: The proposed metric and its associated software tools are vendor-independent and can be used to analyze any LCD scanner performance. Furthermore, the method employed can be used in conjunction with the relationships established in this study between LCD and tissue type to extend these concepts to patients' clinical CT images.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app