TU-H-206-02: Novel Linearly-Filled Derenzo PET Phantom Design.

Medical Physics 2016 June
PURPOSE: To design a linearly-filled Derenzo positron emission tomography (PET) phantom, eliminating the extraneous radioisotope volumes in a conventional reservoir-type design. This activity reduction combined with the elimination of bubbles in smaller phantom channels would significantly reduce personnel dose, radioisotope cost, and would improve image quality by reducing out-of-slice activity scatter.

METHODS: A computer-aided design (CAD) was created of a modular Derenzo phantom consisting of three phantom layers with gaskets between the layers. The central piece contains the active pattern volume and channels connecting adjacent rods in a serpentine pattern. The two end-pieces contained an inlet and an outlet for filling purposes. Phantom prototypes were 3D printed on a Viper Si2 stereolithography machine. The two gaskets were fabricated from silicon sheets using a PLS 6.75 laser cutter. Phantoms were held together by pass-through glass-filled nylon bolts and nuts. Phantoms were filled with (52) Mn, (64) Cu, (74) Br, and (124) I for testing, and were imaged on a Siemens Inveon MicroPET scanner.

RESULTS: Four phantom prototypes were constructed using male Leur Lock fittings for inlet/outlet ports. 3D printed layers were sanded to ensure proper coupling to the silicon gaskets. The filling volume for each prototype was approximately 2.4 mL. The filling process was found to be rapid, leak-tight, and with minimal back-pressure. PET images were reconstructed by OSEM3D, and axial slices along the phantom pattern length were averaged to provide final images. Image distortion was isotope dependent with (52) Mn and (64) Cu having the least distortion and (124) I having the most distortion.

CONCLUSION: These results indicate that the linearlyfilled Derenzo design improves on conventional reservoir-type designs by eliminating potential bubbles in small channels and by reducing activity level, radioisotope volume, radioisotope cost, personnel dose, filling time, and out-of-slice activity scatter The method described in this abstract has been filed as a patent application to the US Patent and Trade Office by the Wisconsin Alumni Research Foundation (WARF).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app