SU-F-T-284: The Effect of Linear Accelerator Output Variation On the Quality of Patient Specific Rapid Arc Verification Plans.

Medical Physics 2016 June
PURPOSE: The aim of the current study is to investigate the effect of machine output variation on the delivery of the RapidArc verification plans.

METHODS: Three verification plans were generated using Eclipse™ treatment planning system (V11.031) with plan normalization value 100.0%. These plans were delivered on the linear accelerators using ArcCHECK- device, with machine output 1.000 cGy/MU at calibration point. These planned and delivered dose distributions were used as reference plans. Additional plans were created in Eclipse- with normalization values ranging 92.80%-102% to mimic the machine output ranging 1.072cGy/MU-0.980cGy/MU, at the calibration point. These plans were compared against the reference plans using gamma indices (3%, 3mm) and (2%, 2mm). Calculated gammas were studied for its dependence on machine output. Plans were considered passed if 90% of the points satisfy the defined gamma criteria.

RESULTS: The gamma index (3%, 3mm) was insensitive to output fluctuation within the output tolerance level (2% of calibration), and showed failures, when the machine output exceeds ≥3%. Gamma (2%, 2mm) was found to be more sensitive to the output variation compared to the gamma (3%, 3mm), and showed failures, when output exceeds ≥1.7%. The variation of the gamma indices with output variability also showed dependence upon the plan parameters (e.g. MLC movement and gantry rotation). The variation of the percentage points passing gamma criteria with output variation followed a non-linear decrease beyond the output tolerance level.

CONCLUSION: Data from the limited plans and output conditions showed that gamma (2%, 2mm) is more sensitive to the output fluctuations compared to Gamma (3%,3mm). Work under progress, including detail data from a large number of plans and a wide range of output conditions, may be able to conclude the quantitative dependence of gammas on machine output, and hence the effect on the quality of delivered rapid arc plans.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app