TU-FG-BRB-01: Dual Energy CT Proton Stopping Power Ratio Calibration and Validation with Animal Tissues.

Medical Physics 2016 June
PURPOSE: The conversion of Hounsfield Unit (HU) to proton stopping power ratio (SPR) is a main source of uncertainty in proton therapy. In this study, the SPRs of animal tissues were measured and compared with prediction from dual energy CT (DECT) and single energy CT (SECT) calibrations.

METHODS: A stoichiometric calibration method for DECT was applied to predict the SPR using CT images acquired at 80 kVp and 140 kVp. The dual energy index was derived based on the HUs of the paired spectral images and used to calculate the SPRs of the materials. Tissue surrogates with known chemical compositions were used for calibration, and animal tissues (pig brain, liver, kidney; veal shank, muscle) were used for validation. The materials were irradiated with proton pencil beams, and SPRs were deduced from the residual proton range measured using a multi-layer ion chamber device. In addition, Gafchromic EBT3 films were used to measure the distal dose profiles after irradiation through the tissue samples and compared with those calculated by the treatment planning system using both DECT and SECT predicted SPRs.

RESULTS: The differences in SPR between DECT prediction and measurement were -0.31±0.36% for bone, 0.47±0.42% for brain, 0.67±0.15% for liver, 0.51±0.52% for kidney, and -0.96±0.15% for muscle. The corresponding results using SECT were 3.1±0.12%, 1.90±0.45%, -0.66±0.11%, 2.33±0.21%, and -1.70±0.17%. In the film measurements, average distances between film and calculated distal dose profiles were 0.35±0.12 mm for DECT calibration and -1.22±0.12 mm for SECT calibration for a beam with a range of 15.79 cm.

CONCLUSION: Our study indicates that DECT is superior to SECT for proton SPR prediction and has the potential to reduce the range uncertainty to less than 2%. DECT may permit the use of tighter distal and proximal range uncertainty margins for treatment, thereby increasing the precision of proton therapy.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app