SU-F-J-183: Interior Region-Of-Interest Tomography by Using Inverse Geometry System.

Medical Physics 2016 June
PURPOSE: The inverse geometry computed tomography (IGCT) composed of multiple source and small size detector has several merits such as reduction of scatter effect and large volumetric imaging within one rotation without cone-beam artifact, compared to conventional cone-beam computed tomography (CBCT). By using this multi-source characteristics, we intend to present a selective and multiple interior region-of-interest (ROI) imaging method by using a designed source on-off sequence of IGCT.

METHODS: All of the IGCT sources are operated one by one sequentially, and each projection in the shape of narrow cone-beam covers its own partial volume of full field of view (FOV) determined from system geometry. Thus, through controlling multi source operation, limited irradiation within ROI is possible and selective radon space data for ROI imaging can be acquired without additional X-ray filtration. With this feature, we designed a source on-off sequence for multi ROI-IGCT imaging, and projections of ROI-IGCT were generated by using the on-off sequence. Multi ROI-IGCT images were reconstructed by using filtered back-projection algorithm. All these imaging process of our study has been performed by utilizing digital phantom and patient CT data. ROI-IGCT images of the phantom were compared to CBCT image and the phantom data for the image quality evaluation.

RESULTS: Image quality of ROI-IGCT was comparable to that of CBCT. However, the distal axial-plane from the FOV center, large cone-angle region, ROI-IGCT showed uniform image quality without significant cone-beam artifact contrary to CBCT.

CONCLUSION: ROI-IGCT showed comparable image quality and has the capability to provide multi ROI image within a rotation. Projection of ROI-IGCT is performed by selective irradiation, hence unnecessary imaging dose to non-interest region can be reduced. In this regard, it seems to be useful for diagnostic or image guidance purpose in radiotherapy such as low dose target localization and patient alignment. This research was supported by the Mid-career Researcher Program through NRF funded by the Ministry of Science, ICT & Future Planning of Korea (NRF-2014R1A2A1A10050270) and by the Radiation Technology R&D program through the National Research Foundation of Korea funded by the Ministry of Science, ICT & Future Planning (No. 2013M2A2A7038291).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app