SU-F-T-172: A Method for Log File QA On An IBA Proteus System for Patient Specific Spot Scanning Quality Assurance.

Medical Physics 2016 June
PURPOSE: The use of log files to perform patient specific quality assurance for both protons and IMRT has been established. Here, we extend that approach to a proprietary log file format and compare our results to measurements in phantom. Our goal was to generate a system that would permit gross errors to be found within 3 fractions until direct measurements. This approach could eventually replace direct measurements.

METHODS: Spot scanning protons pass through multi-wire ionization chambers which provide information about the charge, location, and size of each delivered spot. We have generated a program that calculates the dose in phantom from these log files and compares the measurements with the plan. The program has 3 different spot shape models: single Gaussian, double Gaussian and the ASTROID model. The program was benchmarked across different treatment sites for 23 patients and 74 fields.

RESULTS: The dose calculated from the log files were compared to those generate by the treatment planning system (Raystation). While the dual Gaussian model often gave better agreement, overall, the ASTROID model gave the most consistent results. Using a 5%-3 mm gamma with a 90% passing criteria and excluding doses below 20% of prescription all patient samples passed. However, the degree of agreement of the log file approach was slightly worse than that of the chamber array measurement approach. Operationally, this implies that if the beam passes the log file model, it should pass direct measurement.

CONCLUSION: We have established and benchmarked a model for log file QA in an IBA proteus plus system. The choice of optimal spot model for a given class of patients may be affected by factors such as site, field size, and range shifter and will be investigated further.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app