SU-F-T-260: Using Portal Image Device for Pre-Treatment QA in Volumetric Modulated Arc Plans with Flattening Filter Free (FFF) Beams.

Medical Physics 2016 June
PURPOSE: To implement and validate a method of using electronic portal image device (EPID) for pre-treatment quality assurance (QA) of volumetric modulated arc therapy (VMAT) plans using flattering filter free (FFF) beams for stereotactic body radiotherapy (SBRT).

METHODS: On Varian Edge with 6MV FFF beam, open field (from 2×2 cm to 20×20 cm) EPID images were acquired with 200 monitor unit (MU) at the image device to radiation source distance of 150cm. With 10×10 open field and calibration unit (CU) provided by vendor to EPID image pixel, a dose conversion factor was determined by dividing the center dose calculated from the treatment planning system (TPS) to the corresponding CU readout on the image. Water phantom measured beam profile and the output factors for various field sizes were further correlated to those of EPID images. The dose conversion factor and correction factors were then used for converting the portal images to the planner dose distributions of clinical fields. A total of 28 VMAT fields of 14 SBRT plans (8 lung, 2 prostate, 2 liver and 2 spine) were measured. With 10% low threshold cutoff, the delivered dose distributions were compared to the reference doses calculated in water phantom from the TPS. A gamma index analysis was performed for the comparison in percentage dose difference/distance-to-agreement specifications.

RESULTS: The EPID device has a linear response to the open fields with increasing MU. For the clinical fields, the gamma indices between the converted EPID dose distributions and the TPS calculated 2D dose distributions were 98.7%±1.1%, 94.0%±3.4% and 70.3%±7.7% for the criteria of 3%/3mm, 2%/2mm and 1%/1mm, respectively.

CONCLUSION: Using a portal image device, a high resolution and high accuracy portal dosimerty was achieved for pre-treatment QA verification for SBRT VMAT plans with FFF beams.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app