TH-EF-BRB-06: Implementation of a Modulated-Arc Total Body Irradiation (TBI) Technique Using the RayStation Treatment Planning System.

Medical Physics 2016 June
PURPOSE: To develop a clinical workflow for delivering a modulated-arc total body irradiation (TBI) with RayStation scripting. This technique uses arc fields with the patient lying at floor level on a padded table and is validated through measurements taken on a custom-made TBI phantom.

METHODS: Treatment planning was performed for a retrospective cohort of eight patients with a diverse range of heights and body types. Each was replanned using an open-field dual arc method, with the patient in supine and prone positions on the floor of the vault. All plans were optimized using Raystation Planning 4.7.2.5 (RaySearch Laboratories, Stockholm, Sweden), with 200 cGy prescribed to the 95% of the body contour - 5mm. This results in an open-field beam that sweeps craniocaudally across the length of the patient. The technique is validated with measurements at 10 cm intervals in a custom-milled, 5 cm thick acrylic phantom. A centrally located CC13 ion chamber and a Mobile MOSFET (Best Medical Canada, Ottawa, ON) detector array were used to measure dose. Supine and prone arcs for each patient were consecutively delivered, and the aggregate dose at each point was compared to the planned dose calculated in the phantom.

RESULTS: The ion chamber measurements differed from the planned dose by an average of .5%, with a standard deviation of 2.1%. All measured data for the MOSFETS were within 10% of the corresponding planned dose except for two outlying points. The standard deviation of dose differences across the entire cohort was 4.0%. Most significant discrepancies occurred either in inhomogeneous regions with large gradients, or at inferior points where beam angle was steepest.

CONCLUSION: We have confirmed that the planned dose is well matched to our measurements within 10% for this method of planning and delivery. We are currently incorporating this technique into our clinical workflow. This work is supported by RaySearch.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app