TU-FG-209-06: Quantitative Evaluation of the Temporal Performance of Clinical Fluoroscopic Imaging Systems: The Temporal Modulation Transfer Function (TMTF).

Medical Physics 2016 June
PURPOSE: Measure the temporal modulation transfer function (TMTF) of clinical fluoroscopic flat panel imaging systems in order to accurately quantify their performance for temporally sensitive clinical tasks.

METHODS: Copper blades (0.76 mm thick, 6 cm radius) with precision-machined edges were manufactured and mounted on a voltage regulated DC motor apparatus. Images were acquired with the blade apparatus positioned at the center of the detector matrix and set in motion at a constant rotational velocity (0.66 Hz), thereby creating a rotating radio-opaque edge in both space and time. The spatio-temporal edge response function was analyzed using a single frame from the acquired image sequence. Image processing included semi-automatic detection of the center of rotation, rebinning the pixels into subsampled polar coordinates, and deconvolving the TMTF from the previously measured spatial MTF. The analysis returned the presampled TMTF. This method was applied to multiple fluoroscopic imaging systems.

RESULTS: Initial experiments measured the TMTF of a typical fluoroscopy unit (Philips Allura Xper FD20) using cine acquisition (60.7 kVp, 6 fps, 42.6 ms pulse-width). For this system and protocol, the measured TMTF closely matched an idealized sinc function corresponding to the Fourier transform of the measured rectangular x-ray pulse-width. Within the frequency range of 0 - 47 Hz, the measured TMTF and ideal sinc function were compared at 1.46 Hz intervals. The differences measured varied within [-0.0164, 0.0324] with a root-mean-square (rms) difference of 0.0201. These particular results suggest that very little degradation in temporal performance is attributable to the imaging hardware, but rather the measured performance is dominated by the acquisition protocol parameters, namely pulse-width and frames-per-second.

CONCLUSION: This method provides a clinically tractable and accurate measurement of the TMTF. Ongoing experiments are investigating the impact of different fluoroscopy systems, image processing, and protocol choice on the measured TMTF.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app